2026 年 7 月 29 日,英国药品和健康产品管理局(MHRA)与英格兰 NHS 联合发布环境语音技术产品(AVT,即 AI 病历助手)指引。文件给出 5 类不算医疗器械的功能:只做转录、只做摘要、只整理结构化数据、只按对话明确提到的术语建议临床编码、只起草出院小结或信函草稿。反过来,预期用途指向诊断、治疗、预防,或代替医生自动执行动作(如未复核即下单)的产品,仍按医疗器械监管。指引不改变法律,只解释 2002 年英国医疗器械法规(UK MDR)怎么适用,并要求厂商在每次功能变更后重新判断产品是否落入器械定义。
- 来源机构:英国药品和健康产品管理局(MHRA)新闻稿:MHRA clarifies regulatory status of ambient voice technologies used in the NHS;MHRA 指引:Ambient voice technology-enabled products;MHRA 指引 PDF:Guidance on Ambient Voice Technology-Enabled Products;NHS England:Guidance on the use of AI-enabled ambient scribing products in health and care settings
- 原文发布日期:2026-07-29
- 本站采集时间:2026-09-28T03:56:02.621Z
- 本站发布时间:2026-09-28 11:56
- 文章来源:英国药品和健康产品管理局(MHRA)新闻稿:MHRA clarifies regulatory status of ambient voice technologies used in the NHS;MHRA 指引:Ambient voice technology-enabled products;MHRA 指引 PDF:Guidance on Ambient Voice Technology-Enabled Products;NHS England:Guidance on the use of AI-enabled ambient scribing products in health and care settings
- 英国药品和健康产品管理局(MHRA)新闻稿:MHRA clarifies regulatory status of ambient voice technologies used in the NHS:https://www.gov.uk/government/news/mhra-clarifies-regulatory-status-of-ambient-voice-technologies-used-in-the-nhs(2026-07-29) MHRA 指引:Ambient voice technology-enabled products:https://www.gov.uk/government/publications/ambient-voice-technology-enabled-products/ambient-voice-technology-enabled-products MHRA 指引 PDF:Guidance on Ambient Voice Technology-Enabled Products:https://assets.publishing.service.gov.uk/media/6a68daaff23ff62c51c1a7a3/MHRA_Guidance_on_Ambient_Voice_Technology-Enabled_Products.pdf NHS England:Guidance on the use of AI-enabled ambient scribing products in health and care settings:https://www.england.nhs.uk/long-read/guidance-on-the-use-of-ai-enabled-ambient-scribing-products-in-health-and-care-settings/
完整原文中文翻译
环境语音技术(ambient voice technology,AVT)这两年铺得很快。医生边问诊,AI 边把对话转成文字、写成摘要。问题是:这些东西到底算不算医疗器械。
2026 年 7 月 29 日,英国 MHRA 给出了答案。
一份只回答一个问题的文件
MHRA 当天发布了一份指引,主题是环境语音技术产品(AVT products),也就是市面上常说的 AI 病历助手、AI scribe。文件由 MHRA 与英格兰 NHS 共同编写。
它要回答的问题只有一个:这类产品在什么情况下算医疗器械,在什么情况下不算。
需要先说清楚边界。这不是新法律。它不改变 2002 年英国医疗器械法规(UK MDR)的任何一条。指引写的是 MHRA 如何把现有法规套用到这类产品上。
同一天,NHS England 也发布了配套的临床使用指引。
不算医疗器械的 5 类功能
MHRA 用 5 个例子划出了低风险区。
第一类,只做转录。把医患对话转成文字,交给医生复核、修改。产品不额外给建议。
第二类,只做摘要。把一次就诊对话总结成摘要,供医生复核后再写入电子病历。
第三类,把对话信息整理成结构化数据。例如填进问题清单、用药清单、医嘱集,供医生确认或驳回。
第四类,根据对话里明确说出的术语,建议可能的临床编码。它不推断诊断,只做字面匹配。
第五类,起草出院小结或信函草稿,供医生修改。
这 5 类的共同点是:只碰行政和文档,不产出新的临床判断。
指引里还有一句话值得记住。把产品用在医疗环境里,这件事本身不会让它变成医疗器械。
反过来,这两类仍然算器械
如果产品的预期用途是支持诊断、治疗或预防,它就是医疗器械。
如果产品会代替医生自动执行动作,例如未经医生复核就下单,同样是医疗器械。
那预期用途怎么认定?看标签、说明书,也看宣传材料。厂商官网上的描述、销售演示里的说法、线上广告,都会被算进去。
这里有个真实的坑。指引明确说,一句「本产品不用于诊断」的免责声明,救不了在别处做过的医疗宣称。声明是最低一级的风险控制手段,厂商应优先用设计去消除风险。
每次功能更新都要重新判断
MHRA 提醒厂商:上线时不是器械,不等于以后都不是。
厂商新增功能、扩大用途之后,产品可能就落进医疗器械的定义里。指引写得很直接,每一次变更都要重新评估。
指引还单独点了一个技术风险。大模型的幻觉,在 AVT 产品里影响面很广。厂商要按产品的预期用途去识别和降低这类风险,并且优先改设计,而不是加警告。
对企业与临床机构的意义
对厂商,这是一次口径校准。产品算不算器械,直接决定合规路径。
按 MHRA 独立软件指引的分类规则,用于诊断的有源器械落在 IIa 类,其余有源器械落在 I 类。达到 IIa 类及以上,就需要公告机构出具 UKCA 认证。
适用范围必须讲清楚。这套规则覆盖大不列颠,也就是英格兰、威尔士、苏格兰。北爱尔兰依据温莎框架,继续适用欧盟 MDR(2017/745)。
对临床机构,MHRA 要求 NHS 董事会确认 4 件事:有临床监督、有本地治理、有员工培训、有采购流程。并且要能识别产品功能变化带来的监管地位变化。
医生仍然要复核 AI 生成的转录、摘要和其他输出。这条责任不因指引而改变。
来源与适用说明
① 来源:英国药品和健康产品管理局(MHRA)新闻稿「MHRA clarifies regulatory status of ambient voice technologies used in the NHS」,发布日 2026-07-29;MHRA 指引《Ambient voice technology-enabled products》;MHRA 指引 PDF;NHS England 配套指引《Guidance on the use of AI-enabled ambient scribing products in health and care settings》。
https://www.gov.uk/government/news/mhra-clarifies-regulatory-status-of-ambient-voice-technologies-used-in-the-nhs
https://www.gov.uk/government/publications/ambient-voice-technology-enabled-products/ambient-voice-technology-enabled-products
https://assets.publishing.service.gov.uk/media/6a68daaff23ff62c51c1a7a3/MHRA_Guidance_on_Ambient_Voice_Technology-Enabled_Products.pdf
https://www.england.nhs.uk/long-read/guidance-on-the-use-of-ai-enabled-ambient-scribing-products-in-health-and-care-settings/
② 核验范围与访问日期:2026-09-28 访问并核验上述 MHRA 新闻稿、指引页与指引 PDF 正文(含第 4 节器械资格判断、第 5 节资格举例、Rule 10 与 Rule 12 分类引用),以及 NHS England 配套指引页面。指引内所引 UK MDR 2002 与欧盟 MDR 2017/745 的条款号以官方原文为准。
③ 适用说明:本文是对公开监管信息的解读,不构成对任何具体产品、企业或注册申请的结论与承诺。具体产品是否属于医疗器械,须结合其标签、说明书与宣传材料中的预期用途逐案判断,并以官方原文与后续更新为准。
这条信息对医疗器械企业意味着什么?
AI医疗器械和软件医疗器械的海外注册,监管重点通常不只在产品功能描述,还包括算法用途、风险控制、数据来源、性能验证、网络安全、软件生命周期、临床评价和上市后持续学习或版本变更管理。
不同国家对AI/软件医疗器械的分类、证据要求和变更管理要求不完全一致,企业不能只用传统硬件医疗器械的资料框架来准备。
企业需要重点检查哪些资料?
建议优先整理软件描述、算法预期用途、输入输出、适用人群、训练和验证数据说明、性能验证报告、风险管理、软件生命周期文件、网络安全资料、临床评价资料、标签说明书、版本管理记录和上市后监测计划。
如产品涉及诊断、治疗建议、生命支持、自动决策或高风险场景,还需要更谨慎评估证据深度和临床验证要求。
常见风险
常见问题包括:算法用途描述过宽、训练数据和验证数据边界不清、软件版本变更缺少控制策略、网络安全资料不足、临床证据与宣称不匹配,以及不同国家分类结论不一致。
艾梅莉观点
这条指引对做 AI 医疗软件的企业,影响是直接的:产品定义改一句话,合规成本可能差一个数量级。
先看差别。被判为医疗器械,就要走资质、分类、技术文件、上市后监督这一整套。判为非器械,就只受数据、隐私、采购等常规规则约束。IIa 类及以上还要公告机构认证。
所以企业要做的第一件事,是把自己的产品拆到功能级去看。同一个 App,转录模块可能不算器械,但一旦加入「根据对话推断诊断」或「自动修改医嘱」,整条边界就跨过去了。
第二件事,是清理宣传口径。指引说得很清楚,免责声明不能抵消医疗宣称。官网文案、销售 PPT、演示视频、投放广告,都要一起过一遍。只要有一处暗示诊断或治疗,前面所有「仅供记录」的说明都白写。
第三件事,是把变更评估写进流程。产品每次加功能、换模型、扩场景,都要重新做一次器械资格判断,并把结论留档。这件事不做的代价,通常是事后整改加下架。
第四件事,别忽略幻觉风险。指引把大模型幻觉当成需要按预期用途管理的风险。厂商要在设计阶段做缓解,而不是事后加提示语。
对已在国内或欧盟拿证、又想做英国市场的团队,还有一层衔接要注意。大不列颠用 UKCA,北爱尔兰继续用 CE。两套规则的软件分类不完全一样,注册和标签不能直接照搬。
需要提醒适用边界。这是英国的规则,覆盖大不列颠,不覆盖北爱尔兰,也不代表沙特、巴西、东南亚等市场的口径相同。企业应按目标市场逐地判断,而不是拿一份指引套全球。
我们的建议是三步走。第一步,出产品功能清单,逐条标注是否构成医疗用途。第二步,按结论决定各自的注册路径与证据要求。第三步,把变更判断固化进内部审批。做完这三步,再谈多市场扩张,顺序才不会反。
最后补一句时间判断。这类软件功能迭代按月计,监管口径也在变。把产品定义与证据策略一次写清,比每季度重新对口径要省事得多。对已有英国客户或渠道的团队,这一条尤其要在合同前确认。口径提前对清,后面的开发节奏和注册计划就不用推倒重来。
一句话总结:先定义产品,再定义路径,顺序不能反过来。先定义产品,再定义路径,顺序别反。
还要提醒一点:指引举例中提到,产品若只把对话内容整理成结构化数据、或只按对话里明确出现的术语给出临床编码建议,同样不算医疗器械。判断的关键始终是预期用途,而不是用了哪种模型,也不是部署在哪个科室。
FAQ
AI医疗器械海外注册最难的地方是什么?
难点通常在风险分类、算法证据、临床评价、软件生命周期和上市后变更管理,而不是单纯填写申请表。
AI医疗器械可以先按普通软件资料申报吗?
不建议简单套用普通软件资料。应先确认产品是否属于SaMD、是否有AI辅助决策属性、是否影响诊断或治疗,再设计注册资料。
- 内容作者:艾梅莉法规编辑部
- 专业审核:艾梅莉医疗器械国际注册项目组
- 来源原则:优先参考官方监管机构、国际组织、标准组织和公开法规资料;行业媒体与项目经验仅作辅助判断。
- 适用边界:本文用于前期理解、资料准备和项目规划,不替代目标国家监管机构正式要求、检测结论或法律意见。
需要评估注册路径?
请提供产品名称、预期用途、目标国家和现有资料基础,艾梅莉可协助判断注册路径、资料缺口和合规风险。
联系艾梅莉